DC V | 400mV/4/40/400/600V (Input Impedance 10MΩ) |
|±0.6%rdg±4dgt (400mV/4/40/400V) | |
|±1.0%rdg±4dgt (600V) | |
AC V | 400mV/4/40/400/600V (Input Impedance 10MΩ) |
|±1.6%rdg±4dgt (20~400mV) | |
|±1.3%rdg±4dgt (4/40V) | |
|±1.6%rdg±4dgt (400/600V) | |
DC A | 400/4000µA/40/400mA/4/10A |
|±2.0%rdg±4dgt (400/4000µA) | |
|±1.0%rdg±4dgt (40/400mA) | |
|±1.6%rdg±4dgt (4/10A) | |
AC A | 400/4000µA/40/400mA/4/10A |
|±2.6%rdg±4dgt (400/4000µA) | |
|±2.0%rdg±4dgt (40/400mA/4/10A) | |
Ω | 400Ω/4/40/400kΩ/4/40MΩ |
|±1.0%rdg±4dgt (400Ω/4/40/400kΩ/4MΩ) | |
|±2.0%rdg±4dgt (40MΩ) | |
Continuity buzzer | 400Ω (Buzzer sounds below 70Ω) |
Diode Test | 1.5V Release Voltage:Approx. 0.4mA Test Current |
Capacitance Test | 40/400nF/4/40/100µF |
Frequency | 5.12/51.2/512Hz/5.12/51.2/512kHz/5.12/10MHz |
DUTY | 0.1~99.9% (Pulse width/Pulse period) ±2.5%±5dgt |
Withstand Voltage | AC 3700V / 1min. |
Applicable standards | IEC 61010-1 CAT.III 300V |
IEC 61010-2-031 | |
IEC 61326 | |
Power Source | R6P (1.5V) × 2 |
Dimensions | 155(L) × 75(W) × 33(D)mm |
Weight | Approx. 260g |
Accessories | 7210A (Test leads) |
8919 (Ceramic fuse [10A/600V])× 1 | |
8923 (Fuse [0.5A/600V])× 1 | |
R6P× 2 | |
Instruction Manual |